ARIA® Radiation Oncology Server Requirements - Rack Mounted

System Size Small and Medium Large
Platform: DELL® R510 - (2U) DELL® R810 - (2U)
Processor: (Dual) Intel® Xeon® E5620 2.4Ghz, 12M Cache,Turbo, HT, 1066MHz (Dual) Intel® Xeon® E7540 2.00GHz, 18M cache, 6.40 GT/s QPI, Turbo, HT, 6C
Bus Type: PCI-E PCI-E
Memory (RAM): 24GB Memory (6x4GB), 1333MHz Dual Ranked RDIMMs for 2 Processors, Optimized 32GB Memory (16x2GB), 1066MHz, Dual Ranked RDIMMs for 2 or 4 Processors
Hard Drives:
(Local Storage)
  • 6x146GB 15K RPM Serial-Attach SCSI 3Gbps 3.5” Hot Plug Hard Drive
  • 6x600GB 15K RPM Serial-Attach SCSI 6Gbps 3.5” Hot Plug Hard Drive
  • 3x146GB 15K RPM Serial-Attach SCSI 6Gbps 2.5” Hot Plug Hard Drive.
    Direct Attached Storage Array (DAS):
    External
    None PowerVault MD1220 with 24 HDD capacity - (2U)
    Hard Drives (DAS) None
  • 12x146GB 15K RPM Serial-Attach SCSI 6Gbps 2.5" Hot Plug Hard Drive
  • 8x600GB 10K RPM Serial-Attach SCSI 6Gbps 2.5" Hot Plug Hard Drive
  • Drive Controller: Internal PERC H700 Integrated RAID[1] Controller with 512MB of battery-backed cache with (50/50) read/write cache enabled
    • Configure channel one: 2x146GB (RAID 1); 4x146GB (RAID 5)
    • Configure channel two: 6x600GB (RAID 5 + 1 Global HS)
    Internal PERC H700 Integrated RAID[1] Controller with 512MB of battery-backed cache with (50/50) read/write cache enabled
    • Configure channel one: 3x146GB (RAID 1 + 1HS)

    • DVD operates from SATA CONTROLLER ON BOARD
    External PERC H800 Integrated RAID[1] Controller with 512MB of battery-backed cache with (50/50) read/write cache enabled
    • Configure channel one: 3x146GB (RAID 1 + 1HS); 8x600GB (RAID 5 + 1 HS)
    • Configure channel two: 9x146GB (RAID 5 + 1 HS)
    Optical Drive: External USB Slim optical DVD ROM Internal Slim optical DVD ROM
    Backup Tape: None[2] None[2]
    Network Interface: Dual Two-Port Embedded Broadcom® NetXtreme II™ 5709 Gigabit Ethernet NIC Dual Two-Port Embedded Broadcom® NetXtreme II™ 5709 Gigabit Ethernet NIC
    Redundancy: PSU and Fans PSU and Fans
    Ports: 1 serial, minimum 2 USB 1 serial, minimum 2 USB
    Operating System[4], [5]: ARIA (Version 10.0): Windows® Server 2008 R1 (Standard) with Service Pack 2 (25 client access license or sufficient seat licenses) ARIA (Version 10.0): Windows® Server 2008 R1 (Enterprise) with Service Pack 2 (25 client access license or sufficient seat licenses)
    File System:
    NT File System (NTFS)
    Server
    • Volume C, (OS & APP): 146GB (RAID 1) @ 15K RPM;
    • Volume D, (Database & Database Logs): ~438GB (RAID 5) @ 15K RPM;
    • Volume E, (Image Directory & Database Dumps): ~2.3TB (RAID 5) @ 15K RPM.
    Server
    • Volume C, (OS & APP): 146GB (RAID 1) @ 15K RPM;
    Direct Attached Storage
    • Volume D, (Database): ~1.0TB (RAID 5) @ 15K RPM;
    • Volume E, (Database Logs): 146GB (RAID 1) @ 15K RPM;
    • Volume F, (Image Directory & Database Dumps): ~3.6TB (RAID 5) @ 10K RPM.
    Windows® OS Architecture: 64-bit 64-bit
    Microsoft® .NET Framework: ARIA(10.0): 1.1, 2.0, 3.0, and 3.5 ARIA(10.0): 1.1, 2.0, 3.0, and 3.5

    Notes:

    1. Software RAID is neither recommended nor supported with the ARIA Radiation Oncology Server.
    2. ARIA Radiation Oncology Server does not include a tape backup device package.
      a. An optional external tape backup device package can be purchased, which would operate from a separate drive controller.
    3. Hardware used should be approved based on the current Microsoft® Windows® Catalog Listing.
    4. Windows® Server 2008 R2 is currently not compatible with ARIA Version 10.0
    5. Other compatible Windows® operating systems include: Windows Server 2003 and Windows Server 2003 R2, Standard or Enterprise, SP2, 32-bit only.
    6. The Operating System (build and version) of the ARIA Radiation Oncology Server must match the operating system (build and version) of the HARRP target server.
    7. 1 Gbps network recommended for maximum speed and performance.
    8. CPU identified above is for Intel®'s Nehalem or Westmere (microarchitecture). Each CPU has up to three or four separate memory controller hubs (MCHs) within the CPU package, and requires each memory channel to be populated per CPU for optimal use.
    9. Specifications subject to change without notice as newer computer models become available.
    42U Full Height Rack systems come with the following components (see notes below):
    • Cables for Keyboard and Video
    • NEMA standard power strip
    • 1U 8-port keyboard/monitor KVM switch box
    • 1U keyboard/trackball and monitor console
    • 2U 3000VA Uninterruptable Power Supply (UPS)
    • 24-port patch panel (CAT5E)

     

    Contact Varian Oncology

    Varian Oncology Headquarters
    Tel: 1.650.424.5700
    Contact Us

    EMEA

    Customers in the EMEA region, please check EMEA recommended specifications.

    Events

    • 02/02/2012 - 20/04/2012
      ASCO -2012 Genitourinary Cancers Symposium
    • 02/09/2012 - 02/10/2012
      Bard User Meeting - International
    • 02/11/2012 - 02/12/2012
      Singapore Radiological Society 21st ASM
    • 02/21/2012 - 02/23/2012
      HIMSS 2012
    • 02/22/2012 - 02/24/2012
      ACI- Planning, Building & Operating Successful Proton Therpay Center
    • 02/22/2012 - 02/25/2012
      RSS - The Radiosurgery Society
    • 02/23/2012 - 02/25/2012
      ACRO Annual Meeting
    Print Normal Bigger Large

    © 1999-2012 Varian Medical Systems, Inc. All rights reserved.
    Production of any of the material contained herein in any format or media without the express written permission of Varian Medical Systems is prohibited.

    Product Index A - Z